Newmed Hera B18 and B23 B Class Vacuum Autoclave Incl. Printer and USB Logger
DLEX/HERA/B18L
Newmed Hera B18 and B23 B Class Vacuum Autoclave Incl. Printer and USB Logger | VSDent
B18 + USB Logger starting from £ 4,690.00 + VAT
B23 + USB Logger starting from £ 5,090.00 + VAT
Contact us to avail your best package.
Developed with leading technology and designed for high performance, the Hera sterilizer range is innovative, reliable, and easy to use. Available in B or S Class and 18 and 23-litre chamber sizes, Hera is a completely new experience thanks to the blue, backlit capacitive touchpad with acoustic signals to keep the staff informed. The matt finish cover and scratch-free aluminum details give your practice a new look!
Door Locking: Motor-operated door locking guarantees total safety, greater ease of handling and reliable self-regulation of the hatch.
Built in Printer: A standard accessory that during the sterilization process, an automatic print-out records the entire procedure, thereby providing a permanent record of every sterilization cycle. The paper printout can then be kept and attached to your files for a complete and accurate record.
Suitable: Dentistry, Implant / Invasive Dentistry, Medical Sciences
QuikSteam™: With exclusive QuikSteam™ patented technology ensures up to 20% accelerated cycle times, with lower energy consumption
Water Quality Sensor: Our sterilizers are additionally protected by the inclusion of a water quality sensor, as standard. The quality of the water is therefore constantly monitored. The display informs the user whether the water used for sterilization is good.
Power Consumption: 2400 W
IN Compliance: EN 13060 and PED 2014/68/EU
Chamber: |
18 liters made of thick one-piece moulded steel. |
23 liters made of thick one-piece moulded steel. |
Tray Dimensions: |
4 Tray System as Standard 185W x 285D x 18H (mm) |
4 Tray System as Standard 185W x 440D x 18H (mm) |
Chamber Capacity: |
18 L |
23 L |
Weight: |
52 Kg (without accessories) |
58 Kg (without accessories) |
External Dimensions: |
505W x 615D x 400H (mm) |
505W x 690D x 400H (mm) |
All units come as standard with all ancillary accessories including the tray holder, trays, removal tool, cleaning sponge, and water connections included. We also ensure a low cost of ownership by offering you remarkably affordable service, maintenance options and quality spare parts.
Type ‘B’ Class Cycles
- Type B (vacuum). Incorporates a vacuum stage and is designed to reprocess load types such as hollow, air-retentive and packaged loads. Several different cycles may be provided
- Intended for wrapped solid items (e.g. forceps, dental probes), hollow items (e.g. cannulae, tubing), whether or not they are wrapped, and porous loads e.g. fabrics, swabs, and dressings. Type B cycles must have a drying stage to ensure that the load is dry before the door is opened, which can increase the total cycle time considerably
Additional cost:
- Installation, warranty & liability £624.00 + Vat
The installation will include basic user training, please allow 2 hrs for installation & training. The warranty is 2 years or 1500 cycles, whichever comes sooner.
Additional Services available from Dentalex:
Services in which Dentalex offers:
Servicing, HTM01-05 Validations & PSSR2000 examinations, please see below for current charges:
- Servicing (Annual or every 1000 cycles; whichever comes sooner)
From £312.80 + Vat
- Annual service and validation
From £494.80 + Vat
- PSSR2000 Examinations (alongside servicing)
From £105.00 + Vat
The above services can be arranged through Vsdent.